Class I (most serious) Medical Device Z-2622-2026 Ongoing

Bacteria in the alcohol prep pads inside a surgical kit — found not by a complaint but by a routine sterilization dose audit (Windstone Medical Packaging)

Windstone Medical Packaging, Inc. Reported Jul 8, 2026

The U.S. Food and Drug Administration classified a Windstone Medical Packaging surgical kit as Class I — its most serious category — because the large alcohol prep pads inside it had been recalled by Cardinal Health. That recall followed the discovery of a contaminant, Paenibacillus phoenicis, during a routine sterilization dose audit. The quantity is 45 units.

Recall overview (primary data)

  • ClassificationClass I (most serious)
  • Product typeMedical Device
  • Recalling firmWindstone Medical Packaging, Inc.
  • ReasonKits contain Webcol Large Alcohol Prep Pad which were recalled by Cardinal Health following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.
  • DistributionUS Nationwide distribution in the states of PA, GA, CA, WA.
  • Recall initiated2026-05-21

Key points

  • Recall number Z-2622-2026; FDA Class I, the most serious category; status Ongoing; quantity 45 units.
  • The product is a surgical convenience kit, "Dr. Lewin Pack" (AMS7200D).
  • The reason is that the large alcohol prep pads (Webcol) inside had been recalled by Cardinal Health.
  • That recall followed discovery of the contaminant Paenibacillus phoenicis during a routine sterilization dose audit — not from complaints.
  • Distributed to Pennsylvania, Georgia, California, and Washington. Initiated 2026-05-21; reported 2026-07-08.
  • Of the 1,408 FDA recall records this site holds as of 2026-09-01, complaint appears in 51 reasons, internal in 34, monitoring in 14 and audit in four.

1What a sterilization dose audit actually does

Radiation sterilization is run against a dose established in advance as sufficient to reduce microorganisms to the required level. But the microbial load already present in raw materials and the manufacturing environment — the bioburden — is not constant, so whether the established dose is still adequate has to be confirmed periodically.

That is the dose audit: samples are pulled, their microbial state measured, and the setting validated. It is not an investigation triggered by a failure; it is a procedure for confirming that no failure exists.

2Found by an audit rather than a complaint

The record states that this contamination was found in that periodic audit rather than through complaints from patients or clinicians. Most recalls begin after reports of failures accumulate. Contamination surfaced by an audit means it was caught before harm was reported. Product had reached the market — that is why a recall followed — but the route of discovery marks this as a case where the quality system worked.

3Alcohol prep pads, a thing assumed to be sterile

Disposable alcohol prep pads like Webcol are used to wipe skin before an injection or a blood draw. A tool whose purpose is preventing infection being contaminated itself is the same shape as a skin antiseptic failing sterility, which this site holds as a separate recall.

The quantity here is 45 units and distribution covers four states, but that is the count of finished kits — a different figure from the scale of the upstream pad recall.

4What brought it to light

Of the 1,408 FDA recall records this site holds as of 2026-09-01, counting the words in their reasons that point to how a problem surfaced gives complaint in 51, internal in 34, monitoring in 14 and audit in four.

Complaint51 / 1408
Internal testing or checks34 / 1408
Monitoring14 / 1408
Audit4 / 1408

Few records write down how the problem came to light at all. Among those that do, complaints lead, because a report from someone using the product is the readiest starting point. A case like this one, caught by a routine audit while nobody had run into trouble, appears just four times — an exceptional way to be found, as the records have it.

Why it matters

Contamination caught by a periodic audit reaches downstream companies as information they could not have observed themselves. Whether a route exists for upstream quality results to arrive decides how fast a recall can move. And because the finished-product quantity does not represent the scale of the upstream component, judging the size of an event from downstream numbers misreads it. Being able to query stock and shipments at the component level is the practical requirement.

FAQ

What is a sterilization dose audit?
A periodic procedure confirming that the radiation dose established for sterilization is still adequate. Samples are pulled, their microbial state measured, and the setting validated — done to confirm that no failure exists rather than because one occurred.
What difference does it make that an audit found it?
It means the contamination was caught before reports of harm accumulated. Product had reached the market, but as a route of discovery it marks the quality system working.
Is 45 units the scale of the contamination?
No. Forty-five is the count of finished kits, a different figure from the scale of the upstream alcohol pad recall itself.

Sources (primary)

Source: openFDA (U.S. FDA, CC0 public domain). Data is provided as-is and must not be used for medical decisions. Verify the latest and exact details with the official FDA recall information. This site is not endorsed or certified by the U.S. FDA.

#FDA#Recall#Class I#Medical device#Sterilization#Quality audit#Supply chain
Disclaimer: This site independently summarizes and classifies information based on official data sources. Always verify the latest and accurate information with the official sources. Content on finance, health, legal, and security is information, not advice. This site is not an official website of the U.S. government.